產品說明0

Fluke 712 RTD Calibrator
A handheld single function unit in the familiar Fluke 80 Series DMM package. These calibrators are compact and lightweight with easy-to-use pushbutton interface.
Clear 5 digit LCD display
Measure temperature from RTD output
Simulate RTD output
Operate with 7 types of RTD: 100 200 500 1000;JIS and Minco 3916; Ni 120
Measure additional RTDs using Ohms function
Simulate additional RTDs using Ohms source function
°F or °C selectable
4 shrouded banana jacks
Operating temperature -10 to +55°C
| Temperature range | -10 → +55 °C | | |
| Function | Range | Resolution | Accuracy |
| Resistance | 15Ω to 3200Ω | 0.1Ω | ±0.1Ω |
| RTD measure | 200 to 800°C | | |
| | (Pt 100) | 0.1°C, 0.1°F | ±0.1Ω |
| RTD simulate | 200 to 800°C | | |
| | (Pt 100) | 0.1°C, 0.1°F | ±0.1Ω |
| | | | (0.5 to 2mA) |
| | | | ±0.15Ω |
| Dimensions (L x W x D) | 201 x 98 x 52mm | | |
| Weight | 600g | | (0.15 to 0.5mA) |
規格
| Absolute Maximum Resistance Measurement | 3200Ω |
| Absolute Minimum Resistance Measurement | 0Ω |
| Best Accuracy | ±0.01 % |
| Dimensions | 187 x 87 x 32mm |
| Height | 32mm |
| Interface Type | Push Button |
| Length | 187mm |
| Maximum Operating Temperature | +55°C |
| Maximum Temperature Measurement | +800 °C |
| Minimum Operating Temperature | -10°C |
| Minimum Temperature Measurement | +200°C |
| Model Number p | Fluke 712 |
| Resistance Measurement Resolution | 0.1Ω |
| Weight | 330g |
| Width | 87mm |