產品說明0

9100S Drywell
Simple and convenient to use; drywell has a range to 375°C (707°F) and is perfect for checking/calibrating RTDs, thermocouples, and small bimetal thermometers in the field.
Features rapid heating and cooling, prop stand and RS-232 interface capability
Built-in programmable features include temperature scan rate control, eight set-point memory and adjustable readout in °C or °F
Calibration block has 6 probe holes: 1/16, 1/8, 5/32, 3/16 (x2) and 1/4 in diameter
Range | 35°C to 375°C (95°F to 707°F) |
Accuracy | ±0.25°C at 50°C
±0.25°C at 100°C
±0.5°C at 375°C |
Stability | "±0.07°C at 50°C
±0.1°C at 100°C
±0.3°C at 375°C" |
Resolution | 0.1°C or °F |
Hesting times | 35 to 375°C: 9.5 minutes |
Cooling times | 375 to 100°C: 14 minutes |
Well depth | 4 inches (102 mm)
1/16 inch (1.6 mm) hole is 3.5 inches (89 mm) deep |
Power | 115 VAC (±10%), 50//60 Hz, 1.5 A or 230 VAC (±10%), 0.8 A, 50//60 Hz, 175 W |
Fault protection | Sensor burnout protection, over-temperature cutout, and electrical fuses |
Dimensions | 57 (H) x 125 (W) x 150 (D) mm |
Weight | 1.08 kg |
規格
Accuracy | 0.25 °C |
Dimensions | 150 x 125 x 57mm |
Height | 57mm |
Length | 150mm |
Maximum Operating Temperature | +50°C |
Maximum Source Current | 1.5A |
Maximum Source Frequency | 65Hz |
Maximum Source Voltage | 230V ac |
Minimum Operating Temperature | +5°C |
Minimum Source Current | 0.8A |
Minimum Source Frequency | 45Hz |
Minimum Source Voltage | 115V ac |
Model Number p | 9100S |
Stability | ±0.07°C |
Thermocouple Type | RTD |
Weight | 1.08kg |
Well to Well Conformity | ±0.2°C |
Width | 125mm |